EN
LSST-04
LSST-04
LSST-04
LSST-04
LSST-04
LSST-04

Catheter Burst Pressure Tester

LSST-04

LSST-04 Catheter Burst Pressure Tester, adopting the positive pressure test principle, is applicable to the determination of burst pressure under static conditions, rated burst pressure (RBP), balloon fatigue for freedom from leakage and damage on inflation, balloon deflation time, and the relationship between the balloon diameter and the inflation pressure.

Product Features

Embedded high-speed microcomputer chip control

 

Provide multiple test modes for different test items, suitable for balloon and catheter testing

 

Pressure stability and pressure maintaining system conform to test requirements

 

Fluid-filled test chamber, providing 37℃ temperature control environment

 

Adopt self-developed high-end servo booster system, no need to connect to air source, stable test

 

With standard Luer connector, can test various catheters

 

Built-in high-pressure resistant solenoid valve, high-pressure stainless steel pipeline, to ensure the provision of a high-pressure test environment

 

Connection pipelines and screws are made of corrosion-resistant and stainless steel materials, and pipelines are not easy to rust

 

Safety designs such as overpressure safety protection and power-off memory ensure system stability during testing

 

Equipped with standard RS232 interface, which is convenient to PC connection and data transmission

 

Originality Design

 

7-inch HD color LCD screen, real-time display of test data and curves

 

Automatic calibration, simple operation process

 

With a powerful data storage function, which can store up to 100,000 test records

 

Automatic result determination, and printable test data

 

Safety protection

 

With shutdown function for pressure over-range, which protects the sensor.

 

Equipped with emergency stop switch to ensure safe testing

 

Data security

 

Equipped with a four-level user security password policy to ensure the security of test data

 

The computer software fully complies with the requirements of GMP "Computerized System" and has functions such as user management, authority management, and data audit tracking.

Reference Standard

  • ISO10555-1
  • ISO10555-4
  • YY0285.1
  • YY0285.4

Product Parameter

Specifications

LSST-04

Test Range

0~10Mpa

Accuracy

0.5% FS

Connector

Luer taper

Test Mode

1. Test for burst pressure under static conditions2. Test for rated burst pressure (RBP)3. Balloon fatigue test for freedom from leakage and damage on inflation4. Test for balloon deflation time5. Test for balloon diameter to inflation pressure

Constant Temperature Tank

RT+5~100 ℃

Temperature Control Accuracy

±0.5℃

Power Supply

AC 220V 50Hz/110V 60Hz

Dimensions

Instrument: 540mm(L)×400mm(W)×425mm(H)Water tank: 285mm(L)×500mm(W)×340mm(H)Booster pump: 280mm(L)×265mm(W)×750mm(H)

Net Weight

Main engine: 31kgLiquid booster pump: 33kg

 

Applications

 

Basic Application

Test for burst pressure under static conditions

Sterile and single-use catheters

Test for rated burst pressure (RBP)

Balloon dilatation catheters

Balloon fatigue test for freedom from leakage and damage on inflation

Test for balloon deflation time

Test for balloon diameter to inflation pressure

 

Configuration

 

Standard Configuration:Instrument, 7-inch touch screen, Booster pump, Constant temperature water tank, Liquid reservoir and Micro printer.

Related products

Need support or have technical questions?

Pubtester Instruments Co., Ltd testing instruments and packaging testing services help you lift efficiency and guarantee product quality. Tap the button to reach our experts and submit your inquiry today.